Cervical Spine with Contrast for Intramedullary Disease |
Required Sequences | Category A: Pulse Sequence and Image Contrast | Category B: Anatomic Coverage and Imaging Planes | Category C: Spatial Resolution |
Sagittal dark fluid | Should not have non-anatomic heterogeneous signal intensity of the cordCSF must be hypointense to the cord/nerve roots so that cord/nerve roots are well-definedShould not be fat suppressed. The fat must be bright, but not so intense that it masks the fat/muscle planeMust show good contrast between the cord and CSFT1 FLAIR is acceptable for this sequence
| Must cover foramen magnum to T1Must cover laterally through the neural foramina
| Slice thickness ≤ 3.0 mmGap ≤ 1.0 mmPixel area ≤ 1.0 mm2
|
Sagittal bright fluid | Should not have non-anatomic heterogeneous signal intensity of the cordCSF must be hyperintense to the cord/nerve roots so that cord/nerve roots are well-definedFat must not be so intense that it masks the fat/muscle planeMust show good contrast between the cord and CSF
| Must cover foramen magnum to T1Must cover laterally through the neural foramina
| Slice thickness ≤ 3.0 mmGap ≤ 1.0 mmPixel area ≤ 1.0 mm2
|
Axial bright fluid | Should not have non-anatomic heterogeneous signal intensity of the cordCSF must be hyperintense to the cord/nerve roots so that cord/nerve roots are well-definedFat must not be so intense that it masks the fat/muscle planeMust show good contrast between the grey and white matter
| Must cover contiguously from C3 to T1 (from C3 inferior endplate to T1 superior endplate)Axial or oblique axials are acceptable
| Slice thickness ≤ 3.0 mmGap ≤ 1.0 mmPixel area ≤ 1.0 mm2
|
Axial dark fluid post contrast | Should not have non-anatomic heterogeneous signal intensity of the cordCSF must be hypointense to the cord/nerve roots so that cord/nerve roots are clearly-definedShould not be fat suppressed, the fat must not be so intense that it masks the fat/muscle planeMust show good contrast between the cord and CSF
| Must cover contiguously from foramen magnum to T1Axial or oblique axials are acceptable
| Slice thickness ≤ 3 mmGap ≤ 1.0 mmPixel area ≤ 1.0 mm2
|
Sagittal dark fluid post contrast | Should not have non-anatomic heterogeneous signal intensity of the cordCSF must be hypointense to the cord/nerve roots so that cord/nerve roots are clearly-definedShould not be fat suppressed, the fat must not be so intense that it masks the fat/muscle planeMust show good contrast between the cord and CSF
| Must cover foramen magnum to T1Must cover laterally through the neural foramina
| Slice thickness ≤ 3 mmGap ≤ 1.0 mmPixel area ≤ 1.0 mm2
|