Revision History
In addition to the exam-specific instructions in the following tables, review the following information prior to examination selection and submission. Failure to follow the guidance below may result in failure of the submitted examination.
Failure to meet anatomic coverage and imaging plane specifications may result in failure.
MRA Abdomen for Renal Artery Stenosis or Vasculitis | |||
If time-resolved MRA (e.g. TRICKS, TWIST) is submitted, only 2 contrast-enhanced series should be submitted, which must include:
| |||
Required Sequences | Category A:Pulse Sequence and Image Contrast | Category B:Anatomic Coverage and Imaging Planes | Category C:Spatial Resolution |
3D contrast-enhanced source images or 3D non-contrast MRA renal images |
| Must cover:
|
|
3D reformatted images (Angiographic Images) |
| Must display:
| N/A |
MRA High Resolution Arch/Carotid Contrast-Enhanced | |||
If time-resolved MRA (e.g. TRICKS, TWIST) is submitted, only 2 contrast-enhanced series should be submitted, which must include:
| |||
Required Sequences | Category A:Pulse Sequence and Image Contrast | Category B:Anatomic Coverage and Imaging Planes | Category C:Spatial Resolution |
3D contrast-enhanced source images |
| Must cover:
|
|
Reformatted contrast-enhanced angiographic images |
Must have good artery to background contrastMust have uniform artery signalMust have minimal to no venous enhancement | Must display in multiple views and should be derived from the contrast MRA sequence:
| N/A |
MRA Carotid Unenhanced | |||
Required Sequences | Category A:Pulse Sequence and Image Contrast | Category B:Anatomic Coverage and Imaging Planes | Category C:Spatial Resolution |
3D time of flight source images |
| Must cover:
|
|
3D time of flight reformatted images (angiographic images) |
Must have good artery to background contrastMust have uniform artery signalMust have minimal to no venous signal |
| N/A |
2D time of flight source images |
| Must cover:
|
|
2D time of flight reformatted images (angiographic images) |
|
| N/A |
MRA Thoracic Aorta | |||
If time-resolved MRA (e.g. TRICKS, TWIST) is submitted, only 2 contrast-enhanced series should be submitted, which must include:
| |||
Required Sequences | Category A:Pulse Sequence and Image Contrast | Category B:Anatomic Coverage and Imaging Planes | Category C:Spatial Resolution |
Non-contrast black or bright blood source images |
| Must cover:
|
|
3D contrast enhanced source images |
Must have good artery to background contrastMust have uniform artery signalMust have minimal to no venous enhancementMust have correct bolus timing | Must cover:
|
|
3D contrast enhanced reformatted images (angiographic images) |
| Must display in multiple views:
| N/A |