Revision History
Introduction
The Nuclear Medicine Accreditation Physics Subcommittee has defined scoring criteria for planar (uniformity and resolution) and SPECT (uniformity, spatial resolution, and contrast). These criteria are based on results obtained from a variety of cameras that are operating satisfactorily.
Data must be collected, and images prepared, according to the phantom instructions. If sites do not follow the testing requirements as specified in these instructions, then sites may be required to resubmit images or may be subject to failing the submission. Note: Specialty scanners may not be capable of meeting the image acquisition parameters, and guidance is provided within these instructions. Contact the ACR for additional questions or clarifications.
Breast imaging units should follow the small FOV requirements; for spatial resolution the four-quadrant bar phantom should be placed on a 45-degree angle to visualize all quadrants.
Common reasons for image rejection:
Planar images acquired with incorrect parameters (e.g., matrix and counts)
Data form does not accurately reflect information for images submitted
Planar images submitted are not acquired as static planar images (Sites should not submit the daily QC results page)
Planar uniformity images with graphic overlays
Screenshot images with the upper threshold not equal to 100% and lower threshold not between 0-5%
Attenuation correction was not applied (required for cameras that rotate 360 degrees; for SPECT CT units, CT attenuation correction is acceptable)
As appropriate, planar and/or SPECT phantom images must be obtained and submitted for review using the phantom that has been approved by the ACR Committee on Nuclear Medicine Accreditation. Please see the article on Modalities with Low Volume or Emergency Use as needed.
NM QC should be carried out in accordance with the methods outlined in the 2024 NM Quality Control Manual.
The ACR-approved SPECT phantom is commonly used for quality control in nuclear medicine. For cameras that are used to perform planar and SPECT imaging studies, an ACR-approved phantom should be used for evaluating planar and tomographic image quality. The phantom is a cylinder with an internal height of 18.6 cm and internal diameter of 20.3 cm (flangeless phantom). The lower portion of the phantom contains 6 sets of acrylic rods arranged in a pie-shaped pattern with the following diameters: 4.8, 6.4, 7.9, 9.5, 11.1, and 12.7 mm. The upper section contains six solid spheres with the following diameters: 9.5, 12.7, 15.9, 19.1, 25.4, and 31.8 mm.
The ACR has also approved a Small SPECT phantom. The phantom is a cylinder with an internal length of 15 cm and an internal diameter of 13.9 cm (flangeless phantom). The lower portion of the phantom contains 6 sets of acrylic rods arranged in a pie-shaped pattern with the following diameters: 4.8, 6.4, 7.9, 9.5, 11.1, and 12.7 mm. The upper section contains six solid spheres with the following diameters: 6.4, 9.5, 12.7, 15.9, 19.1, and 25.4 mm.
The spatial resolution test can be completed using a four-quadrant bar phantom or an ACR approved SPECT phantom. If a four-quadrant bar phantom is submitted, the smallest bars must be between 2 and 3mm.
Currently the only ACR approved SPECT phantoms are the Data Spectrum Deluxe flanged or flangeless phantom and the Data Spectrum Small flangeless Jaszczak phantom (Small SPECT phantom). The Small SPECT phantom is only approved for use on certain cameras. Submission of accreditation data using any other phantom without ACR permission will disqualify the site.
Cameras that should use Small SPECT phantom: GE 530c, GE 570c, MyoSPECT, maiCam, C!, P3000, and ClearVision
Optional (can use either the Deluxe or Small ACR phantom): c.cam, CardioMD, D-SPECT, Digirad dedicated cardiac SPECT cameras (e.g., Cardius, 2XPO, 3XPO, X-ACT, 2020TC SPECT, SPECTpak), CorCam, and Ventri
The following are available directly from Data Spectrum of Durham, NC (order forms for the large and small phantoms are available in our Testing Overview: Nuclear Medicine and PET article):
Jaszczak Deluxe Flangeless ECT phantom and PET faceplate (can be used for both SPECT and PET acquisitions)
Jaszczak Deluxe Flanged or Flangeless ECT phantom (for SPECT only)
Small SPECT Phantom (for SPECT only)
Flangeless PET phantom (for PET only)
PET faceplate made to fit an existing flangeless or flanged Jaszczak Deluxe ECT phantom
One set of Site Scanning Data Forms (attached below), along with all pertinent worksheets, should be completed for each camera system that is undergoing accreditation.
Phantom Requirements by Module
Module 1 (Planar Imaging Only): Sites that perform only planar imaging on one or more Nuclear Medicine camera systems, and are only applying for accreditation in Module 1 - Planar Imaging on those systems must perform, evaluate, and submit for review images documenting the planar uniformity and spatial resolution for those systems.
All sites applying for Module 1 accreditation are required to submit Tc-99m or Co-57 intrinsic or system planar uniformity images and planar spatial resolution images.
Sites also seeking accreditation for Ga-67 or In-111 must submit Ga-67 or In-111 planar uniformity images and planar spatial resolution images. Sites that perform less than 25 exams using Ga-67 or In-111 in a year are not required to apply for those isotopes.
The Planar protocol should be followed for camera systems only seeking accreditation for Module 1. Images must be submitted for all detectors on systems with more than one detector.
The spatial resolution test can be completed using a four-quadrant bar phantom or the ACR SPECT phantom. If a four-quadrant bar phantom is submitted, the smallest bars must be between 2 and 3mm.
Planar Protocol for Systems Applying for Module 1 (Planar Imaging Only) Accreditation | |
Isotope One: Tc-99m/Co-57 | |
Planar Field Uniformity - Each Detector | Planar Spatial Resolution - Each Detector(Four-Quadrant Bar Phantom or SPECT phantom rods are acceptable) |
Acquire an intrinsic or system uniformity image with the following parameters:
Large rectangular field cameras (smallest dimension ≥ 32 cm), acquire 10 million counts.All circular detectors and small rectangular field cameras (smallest dimension < 32 cm), acquire 5 million counts. | Four Quadrant Bar Phantom: Acquire an intrinsic or system spatial resolution pattern image with the following parameters:
Large rectangular field cameras (smallest dimension ≥ 32 cm), acquire 5 million counts.All circular detectors and small rectangular field cameras (smallest dimension < 32 cm), acquire 3 million counts. |
ACR SPECT Phantom Rods: Fill the ACR phantom with an aqueous solution of Tc-99m, taking care to ensure that it is thoroughly mixed. Place the phantom with the flat bottom on the rods side sitting up and centered on the face of the low-energy collimator.Acquire a static image with the following parameters:
| |
Isotope Two: Ga-67/In-111 | |
Planar Field Uniformity - Each Detector | Planar Spatial Resolution - Each Detector(Four-Quadrant Bar Phantom or SPECT phantom rods are acceptable) |
Acquire an intrinsic or system uniformity image with the following parameters:
Large rectangular field cameras (smallest dimension ≥ 32 cm), acquire 10 million counts.All circular detectors and small rectangular field cameras (smallest dimension < 32 cm), acquire 5 million counts. | Four Quadrant Bar Phantom: Acquire an intrinsic or system spatial resolution pattern image with the following parameters;
Large rectangular field cameras (smallest dimension ≥ 32 cm), acquire 5 million counts.All circular detectors and small rectangular field cameras (smallest dimension < 32 cm), acquire 3 million counts. |
ACR SPECT Phantom Rods: Fill the ACR phantom with an aqueous solution of Ga-67 or In-111, taking care to ensure that it is thoroughly mixed. Place the phantom with the flat bottom on the rods side sitting up and centered on the face of the medium-energy collimator.Acquire a static image with the following parameters:
| |
Planar Images for Submission
| |
Module 2 and/or 3 (SPECT with Planar): Facilities applying for accreditation in the SPECT Module or the Nuclear Cardiology Module must perform, evaluate, and submit for review images documenting the planar uniformity and spatial resolution of each detector, and the SPECT image quality of the system.
All sites applying for Module 2 and/or 3 accreditation are required to submit Tc-99m or Co-57 intrinsic or system planar uniformity images and planar spatial resolution images, except for camera models listed below that cannot acquire planar imaging.
The following cameras are exempt from planar image submission and should proceed to the SPECT Imaging section below: Dedicated cardiac SPECT cameras (e.g., D-SPECT, Digirad, Cardius, 2XPO, 3XPO, X-ACT, 2020TC SPECT, SPECTpak), Veriton, Starguide, GE 530c, GE 570c, MyoSPECT, Clearvision, and GVI.
Sites also seeking accreditation for Ga-67 or In-111 must submit Ga-67 or In-111 planar uniformity images and planar spatial resolution images. Ga-67 or In-111 SPECT phantom images are not required. Facilities that perform less than 25 exams using Ga-67 or In-111 in a year are not required to apply for those isotopes.
The planar and SPECT protocol should be followed for camera systems seeking accreditation for Module 2 and/or Module 3. The spatial resolution test can be completed using a four-quadrant bar phantom or the ACR SPECT phantom. If a four-quadrant bar phantom is submitted, the smallest bars must be between 2 and 3 mm.
Planar and SPECT Protocol for Systems Applying for Module 2 and/or 3 (SPECT and/or Nuclear Cardiology) Accreditation | |
Isotope One: Tc-99m/Co-57 | |
Planar Field Uniformity - Each Detector | Planar Spatial Resolution - Each Detector(Four Quadrant Bar Phantom or SPECT phantom rods are acceptable) |
Acquire an intrinsic or system uniformity image with the following parameters:
Large rectangular field cameras (smallest dimension ≥ 32 cm), acquire 10 million counts.All circular detectors and small rectangular field cameras (smallest dimension < 32 cm), acquire 5 million counts. | Four Quadrant Bar Phantom: Acquire an intrinsic or system spatial resolution pattern image with the following parameters;
Large rectangular field cameras (smallest dimension ≥ 32 cm), acquire 5 million counts.
|
ACR SPECT Phantom Rods: Fill the ACR phantom with an aqueous solution of Ga-67 or In-111, taking care to ensure that it is thoroughly mixed. Place the phantom with the flat bottom on the rods side sitting up and centered on the face of the medium-energy collimator.Acquire a static image with the following parameters:
| |
Isotope Two: Ga-67/In-111 | |
Planar Field Uniformity - Each Detector | Planar Spatial Resolution - Each Detector(Four Quadrant Bar Phantom or SPECT phantom rods are acceptable) |
Acquire an intrinsic or system uniformity image with the following parameters:
Large rectangular field cameras (smallest dimension ≥ 32 cm), acquire 10 million counts.All circular detectors and small rectangular field cameras (smallest dimension < 32 cm), acquire 5 million counts. | Four Quadrant Bar Phantom: Acquire an intrinsic or system spatial resolution pattern image with the following parameters;
Large rectangular field cameras (smallest dimension ≥ 32 cm), acquire 5 million counts.All circular detectors and small rectangular field cameras (smallest dimension < 32 cm), acquire 3 million counts. |
ACR SPECT Phantom Rods: Fill the ACR phantom with an aqueous solution of Ga-67 or In-111, taking care to ensure that it is thoroughly mixed. Place the phantom with the flat bottom on the rods side sitting up and centered on the face of the medium-energy collimator.Acquire a static image with the following parameters:
| |
Planar Images for Submission
| |
SPECT Imaging Protocol
| ||||||||||||||||||||
Positioning of ACR Phantom for Tomography:Place the phantom on the patient support (bed, chair, etc.) so that it is positioned lengthwise, corresponding to the primary patient axis. The phantom’s long axis must be parallel to the rotational (z)-axis of the detector system with the phantom level. Phantoms with flanges may require extra support for leveling. The phantom should be positioned in the center of the field of view of the detector(s).Note: For cameras that are only capable of rotating through 180 degrees and have the detectors in a fixed relative angle of 90 degrees, special caution must be exercised in the orientation of the spheres and rods. You may use your cardiac acquisition protocol. Ensure that the phantom configuration is consistent (largest sphere adjacent to largest rods, second-largest sphere adjacent to second-largest rods, etc.). Position the phantom on the imaging table such that the largest sphere will be in approximately the 10 o'clock position in the reconstructed image. When set up properly, the largest sphere will be in the upper left of the phantom when viewed from the foot of the table. If you are imaging from 0 to 180 degrees, the largest sphere should be at the 12 o'clock position. See the images below for guidance.
| ||||||||||||||||||||
Collimator:Use the highest resolution low-energy parallel hole collimator that is routinely used for clinical studies. | ||||||||||||||||||||
Acquisition:
Sample Calculation for SPECT AcquisitionCalculation of the time per view to acquire approximately 32M counts in the SPECT study: If the system you are using shows the count rate prior to starting the SPECT acquisition, use the procedure given below. If the count rate is not shown, just prior to starting the SPECT study, acquire a planar acquisition for 10 seconds (be sure the imaging table is not between the detector you are using and the phantom) and divide the acquired counts by 10 to get the rate in counts per second.Example:
Assume R = 28,000 counts per second. Assume N = 120 views.Round time up to the next whole number if the fraction of a second is greater than 0.5. Round time down to the next whole number if the fraction of a second is less than 0.5.If the system has two detectors, the total acquisition time will be approximately one-half of the time for a single detector system. If the system has three detectors, the total acquisition time will be approximately one-third of the time for a single detector system.Table for the number of counts to acquire when acquiring the first image by counts and the rest by time
| ||||||||||||||||||||
Reconstruction:
| ||||||||||||||||||||
SPECT Images for Submission
| ||||||||||||||||||||
Evaluation of the Images Submitted
Uniformity, noise, and the presence of artifacts are evaluated qualitatively by inspection of planar images and reconstructed tomographic sections. Planar spatial resolution is judged by identifying the smallest bars in a resolution phantom that can be visualized on cameras used for planar imaging only. For tomographic systems, the smallest “cold” rods in the ACR-approved phantom are used. Collimator type (e.g., general purpose, high resolution) is considered in scoring the spatial resolution. The reconstructed transaxial slices containing the "cold" spheres are used to visually evaluate image contrast based on identifying the smallest “cold” sphere seen at high contrast. As with the evaluation of spatial resolution, collimator type is considered when the score for contrast is determined.
Clarified language in the Planar module 1, Planar module 2, and SPECT Imaging Protocol sections
Clarified language in the Planar module 1, Planar module 2, and SPECT Imaging Protocol sections
![]() | Previous: Nuclear Cardiology Module: NM | Next: Clinical Image Testing: PET |


