Request support
Welcome
Login ACRedit Portal

Stereotactic Breast Biopsy International Recognition Requirements

Personnel Qualifications


Interpreting Physician Personnel Qualifications [1]

  • Must meet country’s qualification requirements

  • Only qualified physicians may independently perform stereotactic breast biopsy procedures at facilities. Physician Assistants or Radiologist’s Assistant cannot perform stereotactic breast biopsies without a qualified physician being in the room at the time of the biopsy procedure.

    Interpreting physician’s responsibilities include:

  • Patient selection (including documentation of correlative clinical breast exams)

  • Quality assurance including the medical audit (tracking of the number of biopsies done, cancers found, benign lesions, biopsies needing repeat, and complications)

  • Oversight of all quality control

  • Supervision of the radiologic technologist and the medical physicist

  • Post-biopsy management of the patient

  • A radiologist practicing in an independent setting is also responsible for:

  • Mammographic interpretation

  • Documentation of correlative breast examinations

  • Referring patients to a surgeon for follow-up on certain lesions

  • Radiologists interpreting stereo exams must meet one of the following Continuing Experience requirements:

  • Currently meets Maintenance of Certification (MOC) requirements for the American Board of Radiology (ABR) or the Osteopathic Continuous Certification (OCC) requirements for the American Osteopathic Board of Radiology (AOBR)

  • 36 image-guided breast biopsies in prior 36 months (at least 9 of these must be stereotactic breast biopsies)

  • Other physicians interpreting stereo exams must meet both of the following Continuing Experience requirements:

  • 36 image-guided breast biopsies in the prior 36 months (at least 9 of these must be stereotactic breast biopsies)

  • Evaluate [2] 720 mammograms in the prior 36 months in consultation with qualified physician

  • Physicians who have not maintained continuing experience may requalify by performing 3 hands-on stereotactic breast biopsy procedures under a qualified physician.

  • All physicians interpreting stereo exams must meet one of the following Continuing Education requirements:


Medical Physicist Personnel Qualifications

  • Must meet country’s qualification requirements

  • Must perform 2 stereotactic breast biopsy unit surveys in the prior 24 months (physicists who have not maintained their continuing experience may requalify by performing 1 hands-on stereotactic breast biopsy physics survey under a qualified medical physicist)

  • Must obtain 3 CEUs in stereotactic breast biopsy in the prior 36 months


Technologist Personnel Qualifications

  • Must have 24 hours of CE every 2 years

    • CE is relevant to imaging and the radiologic sciences, patient care

    • CE includes credits pertinent to the technologist’s ACR-accredited clinical practice

 

 

Stereotactic Breast Biopsy Equipment Quality Control (QC)

Documentation of quality control (QC) is required. Detailed instructions for each of the tests listed below are contained in the 1999 ACR Stereotactic Breast Biopsy Quality Control Manual.


Acceptance Testing

Initial performance testing should be performed upon installation of new stereotactic breast biopsy equipment. This testing should be more comprehensive than periodic performance and compliance testing and should be consistent with current acceptance testing practices.


Annual Medical Physicist Survey


Annual Medical Physicist’s System Performance Evaluation

QC Test

Description

Stereotactic Breast Biopsy Unit Assembly

Ensures that the mechanical components of the system are reliable and safe for patient use

Collimation Assessment

Ensures that the x-ray collimation does not allow significant radiation to extend beyond the edges of the image receptor and that the biopsy window aligns with the x-ray field (Collimation Test criteria requires "5 mm on any side” but per MQSA criteria, it is acceptable in upright systems if the collimation results fall within ±2% of the SID.)

Focal Spot Performance and System Limiting Spatial Resolution

Ensures that the focal spot performance is adequate to minimize geometric blur in the image, and that the system-limiting resolution is adequate for the imaging requirements of the procedure

kVp Accuracy and Reproducibility

Ensures that the indicated peak x-ray energy is accurate and reproducible, so that consistent contrast may be maintained

Beam Quality Assessment (Half-Value Layer Measurement)

Ensures that the x-ray beam is sufficiently penetrating to minimize patient dose, but not so penetrating that contrast is reduced

Automatic Exposure Control (AEC) System or Manual Exposure Performance Assessment

Assesses the performance of the system’s AEC or manual techniques regarding appropriate film optical density or detector signal levels over a range of breast thicknesses

Receptor Speed Uniformity

Ensures that intensifying screens are adequately uniform in speed or that the digital detector is adequately uniform across its entire useful area

Breast Entrance Exposure, Average Glandular Dose and Exposure Reproducibility

Ensures that breast radiation doses are adequately low to protect the patient and sufficient to maintain adequate image quality

Image Quality Evaluation

Ensures that image quality is consistently high enough to meet the demands of the procedure

Artifact Evaluation

Detects the presence of artifacts, isolates their sources and ensures that they are eliminated or minimized

Localization Accuracy Test

Ensures the accuracy of the localization system, including needle position, stereo position calculations and the user interface

Quality Control Tests

A QC program must be implemented for all units and should be established with the assistance of a medical physicist. The radiologic technologist must perform the QC tests listed in the table below at the specified minimum frequencies. The medical physicist should identify the person responsible for performing the tests and may choose to increase the frequency of testing based on the facility and usage. If any QC parameter being monitored falls outside of the control limits, corrective action should be taken. A medical physicist should be available to assist in prescribing corrective actions for unresolved problems.


Radiologic Technologist's QC

QC Test

Description

Frequency

Localization Accuracy Test

Verifies system alignment and performance (procedure varies by manufacturer and system type)

Daily before patient exams

Darkroom Cleanliness (N/A if digital used)

Minimizes artifacts on film images by maintaining the cleanest possible conditions in the darkroom

Daily

Processor QC (N/A if digital used)

Ensures consistent performance of the film processor

Daily

Phantom Images

Ensures that film density, contrast, uniformity, and image quality of the x-ray imaging system are optimal

Weekly

Screen Cleanliness (N/A if digital used)

Ensures that cassettes and screens are free of dust and dirt particles that may degrade image quality or mimic calcifications

Weekly

View boxes and Viewing Conditions (if film used)

Ensures that the view boxes and viewing conditions are optimized and maintained at optimal levels

Weekly

Hardcopy Output Quality (if hardcopy produced from digital data)

Ensures that the quality of hardcopy output is consistent over time and matches the gray scales presented on the CRT monitor

Monthly

Visual checklist

Ensures that the mammography x-ray system and, if applicable, the digital imaging system are working properly and that the mechanical rigidity and stability of the system are optimal

Monthly

Analysis of Fixer Retention in Film (N/A if digital used)

Determines the quantity of residual fixer (hypo) in processed film as an indicator of keeping quality

Quarterly

Compression

Ensures that the x-ray imaging system can provide adequate compression in the manual and automatic powered mode

Semiannually

Repeat Analysis

Determines the number and causes of repeated patient exposures and identifies ways to improve efficiency, reduce patient breast dose, and cut costs

Semiannually

Screen-Film Contact (N/A if digital used)

Ensures that optimum contact is maintained between the screen and the film in each cassette

Semiannually

Darkroom Fog (N/A if digital used)

Ensures that darkroom safelights and other light sources inside and outside of the darkroom do not fog film

Semiannually

Zero Alignment Test (if required by manufacturer)

Verifies that zero coordinate is accurate

Before each patient

Additional tests if required by manufacturer


As required by manufacturer

Preventive Maintenance

Preventive maintenance should be scheduled, performed, and documented by a qualified service engineer on a regular basis. Service performed to correct system deficiencies should also be documented and service records maintained by the facility.


Clinical Image Evaluation and Labeling

Facility’s must review 1 typical clinical examination and attest to meeting the ACR clinical image quality and labeling requirements found below.

Exam Requirements

Correlating Mammogram

  • A 2-view mammogram (CC and MLO/ML/LM)

    Performed within 60 days prior to the stereotactic procedure (Do not review post-procedure images)

    Includes the entire breast with the calcifications to be biopsied circled on each image

    The mammogram can be from an outside facility

    2D and 2D synthesized images are acceptable to review, however, individual slices from tomosynthesis are acceptable only in addition to the CC/ML/MLO views

    Additional images (i.e., XCCL, spot view or magnification views) may be reviewed in addition to, but not instead of, the 2-view mammogram

Stereo Pair

  • The images should demonstrate that physicians possess the skills necessary for appropriate needle positioning during these procedures.

  • Although valuable for some calcifications, “target-on-scout” images are not acceptable for

    the recognition because one cannot assess the needle position on 2 views.

  • Vacuum-Suction biopsy probe (or other FDA-Approved Core Biopsy Device

  • Pre-Fire stereo pair demonstrating needle positioning for tissue acquisition; the calcifications must be visible on both stereotactic views.

OR

  • Post-fire stereo pair demonstrating needle positioning for tissue acquisition; the calcifications MUST be visible on both stereotactic views.

  • Gun-Needle Biopsy:
    • A Pre-Fire stereo pair demonstrating needle positioning; the calcifications must be visible on both stereotactic views.

Specimen Image

  • A specimen image demonstrating calcium.

  • Facilities utilizing the Brevera specimen system should select the single specimen core with the most representative calcifications for review. 


Exam Identification and Labeling

Each image should be clearly and permanently labeled (a permanent sticker label is acceptable for film) with the information below:

  • Patient’s first and last names (required)

  • Identification number and/or date of birth (required)

  • Examination date (required)

  • Facility name

  • Facility location (city, state and zip)

  • Designation of left or right breast (required)

  • Annotation of mammographic view (e.g., CC, MLO/ML/LM)

  • Technologist’s identification number or initials


Physician Quality Assurance

Facilities must conduct ongoing medical audits of stereotactically-guided breast biopsy procedures to evaluate and improve performance. At a minimum, the physician should be able to review the number of procedures done by type, the number of cancers diagnosed and the number of complications requiring treatment. 

  • Total number of procedures

  • Total number of cancers found

  • Total number of benign lesions

  • Total number of stereotactic biopsies needing repeat biopsy, categorized by reason and type of biopsy (i.e., CNB, FNAC)
    • Insufficient Sample
      • Total number of cases

      • Number with repeat biopsy performed by core

      • Number with repeat biopsy performed by excision

      • Final pathology results

    • Discordance
      • Total number of cases

      • Number with repeat biopsy performed by core

      • Number with repeat biopsy performed by excision

      • Final pathology results

    • Cellular Atypia, Radial Scar
      • Total number of cellular atypia cases

      • Total number of radial scar cases (CNB only)

      • Number with repeat biopsy performed by core

      • Number with repeat biopsy performed by excision

      • Final pathology results

    • Other
      • Total number of cases

      • Number with repeat biopsy performed by core

      • Number with repeat biopsy performed by excision

      • Final pathology results

  • Complications requiring treatment, categorized by type of biopsy (i.e., CNB, FNAC)
    • Total number

    • Number with hematoma (requiring intervention)

    • Number with infections requiring treatment

    • Number of other complications

Footnotes

[1] Occasional readers who are providing imaging services to and for the practice readers are not required to meet continuing experience requirements.  However, the reads of all occasional readers combined should not exceed 5% of the total volume of reads per practice and per modality. There must be an active written review process in place at the institution for occasional readers based on each institution’s credentialing requirements.

 

[2] In the context of initial and continuing experience, evaluation means review of the mammographic images in direct consultation with a qualified interpreting physician and/or independent review of mammograms with the authenticated mammographic report.

 


Did you find it helpful? Yes No

Send feedback
Sorry we couldn't be helpful. Help us improve this article with your feedback.