Personnel Qualifications
Interpreting Physician Personnel Qualifications [1]
Must meet country’s qualification requirements
Must meet country’s qualification requirement
Interpret 960 mammographic examinations over a 24-month period
15 Category I CME in mammography in a 36-month period
The ACR Practice Guideline for the Performance of Screening and Diagnostic Mammography recommends that diagnostic mammograms be performed under direct supervision. Direct supervision is defined as the physician being present and immediately available to furnish assistance and direction throughout the performance of the procedure. Direct supervision may also be accomplished via telemammography as long as the interpreting physician is immediately available.
Medical Physicist Personnel Qualifications
Must meet country’s qualification requirements
Survey 2 mammography facilities and 6 mammography units over a 24-month period
Obtain15 CME/CEUs in mammography in a 36-month period
Technologist Personnel Qualifications
Must perform 200 mammographic examinations over a 24-month period
15 Category 1 CEUs in mammography in a 36-month period
Mammography Equipment Quality Control (QC)
Documentation of quality control (QC) is required. Detailed instructions for each of the tests listed below are contained in the
Film/Screen units: 1999 ACR Mammography Quality Control Manual.
FFDM/DBT units: 2018 ACR Digital Mammography Quality Control Manual
Mammography Equipment Evaluations (Acceptance Testing)
The medical physicist is required to perform a Mammography Equipment Evaluation when a new unit or processor is installed, a unit/processor is disassembled and reassembled in the same or a new location, or major components of a mammography unit/processor equipment are changed or repaired. This applies to screen-film, FFDM and DBT units and includes previously owned units, recognized units moved from one site to another, and new units replacing previously recgonized units. All problems must be corrected before the new/changed equipment is put into service for examinations or film processing.
Annual Medical Physicist’s Survey
The medical physicist must perform the QC outlined in the 1999 ACR Mammography QC Manual or 2018 ACR Digital Mammography QC Manual at least annually. The medical physicist should provide a written report of the survey’s findings to responsible physician(s) and professional(s) responsible for service of the equipment.
1999 ACR Mammography QC Manual (Screen-Film Units)
Mammographic Unit Assembly Evaluation
Collimation Assessment
Evaluation of Focal Spot Performance
Automatic Exposure Control (AEC) System Performance
Uniformity of Screen Speed
Artifact Evaluation
Phantom Image Quality Evaluation
kVp Accuracy and Reproducibility
Beam Quality (Half-Value Layer) Assessment
Breast Entrance Exposure, Average Glandular Dose and Radiation Output Rate
Viewbox Luminance and Room Illuminance
2018 ACR Digital Mammography QC ManualACR DM Phantom Image Quality
DBT Z Resolution
Spatial Resolution
DBT Volume Coverage
Automatic Exposure Control System Performance
Average Glandular Dose
Unit Checklist
Computed Radiography (if applicable)
Acquisition Workstation Monitor QC
Radiologist Workstation Monitor QC
Film Printer QC (if applicable)
Evaluation of Site's Technologist QC Program
Evaluation of Display Device Technologist QC Program
Manufacturer Calibrations (if applicable)
Collimation Assessment
ACR DM Phantom Image Quality
DBT Z Resolution
Spatial Resolution
DBT Volume Coverage
Automatic Exposure Control System Performance
Average Glandular Dose
Unit Checklist
Computed Radiography (if applicable)
Acquisition Workstation Monitor QC
Radiologist Workstation Monitor QC
Film Printer QC (if applicable)
Evaluation of Site's Technologist QC Program
Evaluation of Display Device Technologist QC Program
Manufacturer Calibrations (if applicable)
Collimation Assessment
Technologist QC Program
A QC program must be implemented for all units and should be established with the assistance of a medical physicist. The radiologic technologist must perform the QC tests listed below at the minimum frequencies specified in the QC Manual being used. The medical physicist should identify the person responsible for performing the tests and may choose to increase the frequency of testing based on the facility and usage. If any QC parameter being monitored falls outside of the control limits, corrective action should be taken. A medical physicist should be available to assist in prescribing corrective actions for unresolved problems.
1999 ACR Mammography QC Manual (Screen-Film Units)
Darkroom cleanliness
Processor QC
Screen cleaning
Mammo phantom imaging
Film-screen contact test
Compression pressure monitored
Repeat analysis
Viewboxes and viewing conditions
Analysis of fixer retention
Visual checklist
2018 ACR Digital Mammography QC Manual
ACR DM Phantom Image Quality
Computed Radiography Cassette Erasure (if applicable)
Compression Thickness Indicator
Visual Checklist
Acquisition Workstation Monitor QC
Radiologist Workstation Monitor QC
Film Printer QC (if applicable)
Viewbox Cleanliness (if applicable)
Facility QC Review
Compression Force
Manufacturer Calibration (if applicable)
Optional - Repeat Analysis
Record Keeping
Must follow country’s record keeping requirements and/or ACR record keeping requirements below:
Current mammograms and records must be kept by the facility:
For at least 5 years, or
For at least 10 years if no additional mammograms of the patient are performed at the facility, or
For FFDM, either the original or lossless compressed digital data or hard-copy films of final interpretation quality must be maintained.
Preventive Maintenance
Preventive maintenance should be scheduled, performed, and documented by a qualified service engineer on a regular basis. Service performed to correct system deficiencies should also be documented and service records maintained by the facility.
Clinical Image Evaluation and Labeling
Facility’s must review 2 (fatty and dense) typical clinical examinations and attest to meeting the ACR clinical image quality and labeling requirements found below.
Exam Requirements
Should evaluate two Clinical images, a fatty case, and a dense case, 2D acquired or 2D synthesized images, as currently allowed.
Should include CC and MLO images of both breasts
Exam Identification and Labeling
Each image should be clearly and permanently labeled (a permanent sticker label is acceptable for film) with the information below:
Name of patient (first and last)
Additional patient identifier (e.g., medical record number or personal identification number; date of birth is less desirable)
Date of examination
Standardized view (MLO, CC) and laterality (R/L) markers
Facility name and location (must include city, and region)
Technologist identification, such as unique initials
Cassette/screen identification (usually an Arabic numeral written or pressed on the screen). This is used to identify screens with artifacts or defects.
Mammography unit identification number (usually a Roman numeral), if more than one unit in the facility.
A patient ID system for hard copy (strongly recommended)
More permanent than stick-on labels
Information reproduces on copy films
ID should fit squarely in its designated space, near the edge of the film
Not acceptable if any information is illegible, does not fit, or is lopsided, causing cut-off of information
Separate date stickers for hard copy (recommended)
Allows for the date to be easily read with overhead light
May be color-coded by year to facilitate sorting of exams
Technical factors
Include target-filter combination, kVp, mAs, exposure time, compression force, compressed breast thickness and degree of obliquity
Mammography Physician Quality Assurance and Medical Audit
Each facility must establish and maintain a mammography medical outcomes audit program to follow up positive mammography assessments and to correlate pathology results with the interpreting physician’s findings. The ACR strongly recommends that facilities use the ACR Breast Imaging Reporting and Data System (BI-RADS®) final assessment codes and terminology for reporting and tracking of mammography outcomes; however, there is no requirement that the mammography facilities currently recognized by the ACR, nor those applying for recognition in the future, purchase any ACR products in order to obtain or maintain their ACR recognition. The BI-RADS® Atlas also contains guidance on monitoring outcomes and conducting a mammography audit.
[1] Occasional readers who are providing imaging services to and for the practice readers are not required to meet continuing experience requirements. However, the reads of all occasional readers combined should not exceed 5% of the total volume of reads per practice and per modality. There must be an active written review process in place at the institution for occasional readers based on each institution’s credentialing requirements.
Footnotes:
[1] Occasional readers who are providing imaging services to and for the practice readers are not required to meet continuing experience requirements. However, the reads of all occasional readers combined should not exceed 5% of the total volume of reads per practice and per modality. There must be an active written review process in place at the institution for occasional readers based on each institution’s credentialing requirements.