Important: All items in bold in the Technique Parameters and Anatomic Coverage sections below are required to successfully achieve accreditation. CTDIvol information is required for each study. Either include a screen save of the dose with the images, or enter the dose as displayed by the scanner in the Clinical Test Image Data Sheet. Please be aware, special attention must be paid to scanner output (CTDIvol) for pediatric studies. The facility will not receive a pass if the radiation dose is perceived to be too high, regardless of whether the rest of the CT study conforms to satisfactory criteria. Upload the primary reconstructed images used for diagnostic interpretations; include the scout(s) and dose report. Do NOT submit the sagittal, coronal and extra thin images unless specifically listed under required series.
Pediatric Chest |
Technique Parameters |
Required Series | Non-enhanced |
Reconstructed Slice Width | ≤ 5 mm |
Reconstruction Algorithm | Standard or high spatial frequency |
CTDIvol | There are no reference values for this examination |
Anatomical Coverage |
Coverage | Lung apex to below the lung bases |