Important: All items in bold in the Technique Parameters and Anatomic Coverage sections below are required to successfully achieve accreditation. CTDIvol information is required for each study. Either include a screen save of the dose with the images, or enter the dose as displayed by the scanner in the Clinical Test Image Data Sheet.
Adult Head |
Technique Parameters |
Required Series | Unenhanced |
Reconstructed Slice Width | ≤ 5 mm |
Reconstruction Algorithm | Standard or equivalent |
Scan FOV | "Head" or equivalent |
CTDIvol | Reference value: 75 mGy; Pass/Fail criteria: 80 mGy (As reported by the scanner based on a 16 cm phantom) |
Anatomical Coverage |
Coverage | Base of the skull through the vertex. Must include entire head with visualization of the entire bony cranium. |
Lens Exposure | Lens exposure should be avoided |
Adult Temporal Bones |
Technique Parameters |
Required Series | Axial and coronal; With multiple detector helical scanners, coronal images may be reconstructed from axial images |
Reconstructed Slice Width | ≤ 2 mm |
Reconstruction Algorithm | Sharp or equivalent high spatial frequency |
Scan FOV | "Head" or equivalent |
CTDIvol | There are no reference values for this examination |
Anatomical Coverage |
Coverage | Axials: Mastoid tip through petrous bone Coronals: Mandibular condyles through mastoid bone |
Display FOV | 10 cm |
Adult C-Spine |
Technique Parameters |
Required Series | Axial with sagittal and coronal reformations |
Reconstructed Slice Width | ≤ 3 mm |
Reconstruction Algorithm | Sharp (or equivalent) and smooth (or equivalent) |
CTDIvol | There are no reference values for this examination
|
Anatomical Coverage |
Coverage | Cranio-cervical junction to the T1 vertebrae |
Display FOV | 14-16 cm |
Revision History for this Article |
Date | Section | Description of Revision(s) |
12-12-19 | All | Article created; FAQs incorporated; No criteria changes |