Important: All items in bold in the Technique Parameters and Anatomic Coverage sections below are required to successfully achieve accreditation. CTDIvol information is required for each study. Either include a screen save of the dose with the images, or enter the dose as displayed by the scanner in the Clinical Test Image Data Sheet. Upload the primary reconstructed images used for diagnostic interpretations; include the scout(s) and dose report. Do NOT submit the sagittal, coronal and extra thin images unless specifically listed under required series.
Adult Shoulder- Coming Soon |
Technique Parameters |
Required Series | Axial with sagittal and coronal reformations |
Reconstructed Slice Width | ≤ 3 mm |
Reconstruction Algorithm | Sharp (or equivalent) and Smooth (or equivalent) |
CTDIvol | There are no reference values for this examination
|
Anatomical Coverage |
Coverage | Crown of shoulder to mid scapulaScan can be volumetric based upon detector length or helical |
Display FOV | 25 cm recommended |
Adult Hip- Coming Soon |
Technique Parameters |
Required Series | Axial with sagittal, coronal and long oblique axis reformations |
Reconstructed Slice Width | ≤ 3 mm |
Reconstruction Algorithm | Sharp (or equivalent) and Smooth (or equivalent) |
CTDIvol | There are no reference values for this examination |
Anatomical Coverage |
Coverage | Mid-pelvis to lesser trochangerScan can be volumetric based upon detector length or helical |
Display FOV | 22 cm recommended |
Adult Knee- Coming Soon |
Technique Parameters |
Required Series | Axial with sagittal and coronal reformations |
Reconstructed Slice Width | ≤ 3 mm |
Reconstruction Algorithm | Sharp (or equivalent) and Smooth (or equivalent) |
| CTDIvol | There are no reference values for this examination |
Anatomical Coverage |
Coverage | Distal thigh to proximal legScan can be volumetric based upon detector length or helical |
Display FOV | 16 cm recommended |
Adult Foot/Ankle- Coming Soon |
Technique Parameters |
Required Series | Axial with sagittal and coronal reformations |
Reconstructed Slice Width | ≤ 3 mm |
Reconstruction Algorithm | Sharp (or equivalent) and Smooth (or equivalent) |
CTDIvol | There are no reference values for this examination |
Anatomical Coverage |
Coverage | Distal leg to plantar foot surfaceScan can be volumetric based upon detector length or helical |
Display FOV | 16 cm recommended |
Revision History for this Article |
Date | Section | Description of Revision(s) |
1-7-2026 |
| New Module and Exam Criteria Coming Soon |