Technique Parameters and Anatomic Coverage: CT - Calcium Score Module (New- 1-7-2026)

Revision History



Important: All items in bold in the Technique Parameters and Anatomic Coverage sections below are required to successfully achieve accreditation. CTDIvol information is required for each study. Either include a screen save of the dose with the images, or enter the dose as displayed by the scanner in the Clinical Test Image Data Sheet. Upload the primary reconstructed images used for diagnostic interpretations; include the scout(s) and dose report. Do NOT submit the sagittal, coronal and extra thin images unless specifically listed under required series.


If your facility performs any other cardiac exams, you must also apply for the cardiac module.


Adult Calcium Score- Coming Soon

Technique Parameters

Required Series

  • Axials - soft tissue

  • Calcium scoring series with plaques labeled according to vessel

  • Dose report

Dose Reduction

120 kVp (100-140 kVp related to BMI is acceptable)

Reconstructed Slice Width

≤ 3 mm

Reconstruction Algorithm

Soft tissue

CTDIvol

  • N/A

DLP

  • Fail > 120 mGy*cm

Gating

Prospective gating required

Anatomical Coverage

Coverage

Scan must include left main at superior margin of scan range to the bottom of the heart including the PDA at the inferior margin of the scan range 

Required MPR's (MIPS only are not an acceptable substitute for the required MPRs)

  • None




Revision History for this Article

Date

Section

Description of Revision(s)

1-7-2026

All

Article created. New module/exam- Coming Soon

Previous: Technique Parameters and Anatomic Coverage: CT - Adult Cardiac ModuleNext: Technique Parameters and Anatomic Coverage: CT - Pediatric Head and Neck Module